HP To Calibrate the Gain for Channel, HP 53181A DC Power Supply Counter Digital Multimeter

Models: 181A

1 254
Download 254 pages 46.13 Kb
Page 129
Image 129

Chapter 2 Service

The HP 53181A Calibration Procedures

To Calibrate the Gain for Channel 1

1Press any one of the arrow keys until CAL: GAIN 1 ? is displayed.

2Press Enter key.

A scrolling message is displayed. Follow the instruction by performing the following steps.

3Connect a BNC tee connector to Channel 1 of the HP 53181A.

4Connect the Counter, dc power supply, and digital multimeter shown in Figure 2-6.

 

HP 53181A

 

DC Power Supply

Counter

Digital Multimeter

+ - -

BNC Tee

Figure 2-6. Gain Calibration Setup

5Adjust the power supply until the digital multimeter + 5.000 V ± 0.001 V.

6Disconnect the digital multimeter from the Counter to prevent noise from being introduced into the measurement.

7Press Enter key.

The Counter momentarily displays CALIBRATING, and then it should display GAIN 1 PASS.

If the fail message is displayed, refer to the troubleshooting section in this chapter.

2

Assembly-Level Service Guide

2-33

Page 129
Image 129
HP manual To Calibrate the Gain for Channel, HP 53181A DC Power Supply Counter Digital Multimeter