Chapter 2 Service

The HP 53131A/132A Calibration Procedures

CAL: TI QUIK? Calibration

1Press any one of the arrow keys until CAL: TI QUIK? is displayed.

2Press Enter key.

A scrolling message with instructions appears. Follow the instruction by performing the following procedure

3Connect the HP 8130A Pulse Generator output to Channel 1 of the Counter as shown in Figure 2-3.

Equipment

HP 8130A Pulse Generator (or equivalent) HP 10503A BNC Cable

HP 8130A

HP53131A/132A Pulse Generator

Counter

Channel 1

Output

Figure 2-3. Calibration Setup for CAL: TI QUIK?

4Set the HP 8130A as follows: PERIOD: 100 ns

WIDTH: 50 ns

High: 0.50 V Low : −0.50 V

Input Mode: Normal

NOTE

Any external timebase connected to the Counter is ignored during this

 

calibration.

 

 

2-20

Assembly-Level Service Guide