2
Chapter 2 Service
The HP 53131A/132A Calibration Procedures
CAL: TI QUIK? Calibration
1Press any one of the arrow keys until CAL: TI QUIK? is displayed.
2Press Enter key.
A scrolling message with instructions appears. Follow the instruction by performing the following procedure
3Connect the HP 8130A Pulse Generator output to Channel 1 of the Counter as shown in Figure
Equipment
HP 8130A Pulse Generator (or equivalent) HP 10503A BNC Cable
HP 8130A
HP53131A/132A Pulse Generator
Counter
Channel 1 | Output |
Figure 2-3. Calibration Setup for CAL: TI QUIK?
4Set the HP 8130A as follows: PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low : −0.50 V
Input Mode: Normal
NOTE | Any external timebase connected to the Counter is ignored during this |
| calibration. |
|
|