Chapter 2 Service
About the HP 53181A Calibration Menu
About the HP 53181A Calibration Menu
If you are familiar with the information provided in this section, proceed to the section titled “The HP 53181A Calibration Procedures” (page
With the exception of the standard timebase oscillator, no adjustments are required for the HP 53181A Counter.
2
Overview of the HP 53181A Calibration Menu
The Calibration menu is accessed by holding the Scale & Offset key and cycling the POWER key. All of the calibration factors and security settings are stored in
A calibration security code feature allows you to enter a security code (electronic key) to prevent accidental or unauthorized calibrations of the Counter. The security code is set to 53181 when the Counter is shipped from the factory. If you forget your security code, you can reset the security code to the
The Calibration menu, illustrated on page
•view the calibration security status of the Counter (CAL SECURE or
CAL UNSECURE)
•initiate the Counter’s calibration routines, which can perform automatic calibration of the voltage offset and gain of the
•secure against calibration by entering the security code while the Counter is unsecured (CODE: )