Chapter 2 Service
The HP 53131A/132A Calibration Procedures
The HP 53131A/132A Calibration Procedures
First Determine the Counter Firmware Revision
Press and hold Recall (Utility) key, then cycle POWER key.
| REV: n is displayed; where “n” represents the numeric firmware revision |
| code. |
| If 3317, 3335, or 3402 is displayed as the firmware revision code, then |
| perform the calibration procedures in Chapter 6, “Backdating.” |
| If the Counter’s firmware revision code is higher than the ones in the |
| previous paragraph, then perform the following procedures in this section. |
|
|
NOTE | After completing the instrument calibration, BE SURE to secure the |
| calibration by performing the procedure in the section titled “To Secure |
| Against Calibration” on page |
|
|
To Calibrate the Offset for Channels 1 and 2
1To access the Calibration menu, power down, and press and hold the Scale & Offset key, then press POWER key.
The first item in the Calibration menu is a message indicating the calibration security status: CAL SECURE or CAL UNSECURE.
The Counter calibration must be unsecured to perform calibration.
If CAL UNSECURE is displayed, the Counter is already unsecured for calibration; go to step 5. However, if CAL SECURE is displayed then perform steps 2 through 4 to unsecure calibration.
2Press Scale & Offset key until CODE: 0 is displayed.
2