Chapter 1 Performance Tests

Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)

Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)

1

This test verifies the Peak Volts accuracy specification of the HP 53131A and HP 53132A Universal Counters.

Equipment

HP 3325B Synthesizer

HP 10100C 50Ω Feedthrough

HP 10503A BNC Cable

Counter Setup

1Cycle the POWER key to preset the Counter.

2Press Other Meas key until VOLT PEAKS 1 is displayed.3Set CHANNEL 1 input to:50Ω/1MΩ, 50Ω

X10 Attenuate, Off 100kHz Filter, Off

4Set CHANNEL 2 input to:

50Ω/1MΩ, 50Ω

X10 Attenuate, Off 100kHz Filter, Off

Assembly-Level Service Guide