Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
This test verifies the Peak Volts accuracy specification of the HP 53131A and HP 53132A Universal Counters.
Equipment
HP 3325B Synthesizer
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Other Meas key until VOLT PEAKS 1 is displayed.
3Set CHANNEL 1 input to:
50Ω/1MΩ, 50Ω
X10 Attenuate, Off 100kHz Filter, Off
4Set CHANNEL 2 input to:
50Ω/1MΩ, 50Ω
X10 Attenuate, Off 100kHz Filter, Off
1