Chapter 8 HP 53181A Specifications
General Information
General Information
Save and Recall: | Up to 20 complete instrument setups may be saved and recalled later. These setups are retained when | ||
| power is removed from the counter. |
|
|
Rack Dimensions (HxWxD): | 88.5 mm x 212.6 mm x 348.3 mm |
|
|
Weight: | 3.5 kg maximum |
|
|
| AC Line Supply | DC Supply (Option 002 Only) |
|
Power Supply Voltage: | 100 to 120 VAC ±10% - 50, 60 or 400 Hz ±10% | 10 to 32 VDC, | |
| 220 to 240 VAC ±10% - 50 or 60 Hz ±10% |
|
|
| AC Line Voltage Selection: Automatic | Option 002 may not be ordered with Option 060 | |
Power Requirements: | 170 VA maximum (30 W typical) | 4A initial inrush at 10 VDC | |
|
| 3A max, once stabilized | |
Operating Environment: | 0° C to 55° C |
|
|
Storage Environment: |
|
| |
Remote Interface: |
|
| |
| SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2 | ||
Remote Programming Language: | |||
Safety: | Designed in compliance with IEC | ||
EMC: |
|
|
|
| Electrostatic Discharge and Fast Transient/Burst Immunity Testing: When the product is operated at | ||
| maximum sensitivity (20 mVrms) and tested with 8kV AD according to | ||
| transients according to IEC | ||
| made during these disturbances. |
|
|
| Radiated Immunity Testing: When the product is operated at maximum sensitivity (20 mVrms) and tested | ||
| at 3 V/m according to IEC | ||
| The | ||
| counter to any terminal or printer with a properly configured DTE connector | ||
| standard interface cable (HP part number 24542G or 24542H). Data is “output only”; the instrument can | ||
| not be programmed via the |
|
|
8
Note on Pin 4:
1 2 3 4 5
6 7 8 9
May be used as either a DTR signal or an indication of measurement
Pin Number | Type | Description |
|
|
|
2 | Input | Receive Data (RxD) (for Xon/Xoff only) |
|
|
|
3 | Output | Transmit Data (TxD) |
|
|
|
4 | Output | Data Terminal Ready (DTR) |
|
| Measurement |
|
|
|
5 | — | Signal Ground |
|
|
|
6 | Input | Data Set Ready (DSR) |
|
|
|
* | All other pins: no connection | |
|
|
|