1
Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
Test 1: Trigger Level (HP 53181A Only)
This test verifies the Trigger Level accuracy of the HP 53181A Frequency Counter.
Equipment
HP 3325B Synthesizer
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input condition to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000V 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off 100kHz Filter, Off