Chapter 2 Service
Troubleshooting the Counter
|
| Troubleshooting the Counter |
|
| Power Supply Check |
|
| HAZARDOUS VOLTAGES ARE ON THE POWER SUPPLY |
WARNING |
| |
|
| ASSEMBLY. ONLY SERVICE TRAINED AND QUALIFIED |
|
| PERSONNEL SHOULD PERFORM THE FOLLOWING |
|
| PROCEDURE. |
|
|
|
| 1 Remove the power cord from the back of the instrument to | |
|
| disconnect the Counter from the power source. |
| 2 Remove the cover as described in Chapter 3. | |
|
|
|
NOTE |
| A4 AC Power Supply Assembly is fused for safety reasons. It is assumed |
|
| that if the fuses are blown, then damage has occurred to the power supply |
|
| which will make it unreliable. Do not attempt to repair the power supply. |
|
|
|
| 3 Connect the Counter to the power source. | |
|
| The fan should begin to operate as soon as power is supplied. If the fan |
|
| does not operate, replace A4 AC Power Supply Assembly. |
|
| Refer to Chapter 3, “Replacing Assemblies,” in this service guide if you |
|
| need to remove or replace an assembly. |
|
| After replacing the power supply assembly, perform the appropriate |
|
| calibration procedures in this chapter. |
| 4 Connect the negative lead of a voltmeter to the chassis and verify | |
|
| that the voltages in Table |
|
| connector J18 of A1 Motherboard Assembly. See Figure |
|
| illustration of J18 viewing from the bottom of the Counter. |
|
|
|
CAUTION |
| BE CAREFUL to NOT short the |
|
| chassis when probing at the testpoints shown in Figure |
|
|
|
2