How to Calibrate

Calibrate adjusts C-Pen’s reading ability according to the contrast between the text and its background. When the C- Pen display says “Press scan to calibrate”, follow these instructions:

1.Hold C-Pen with the pen tip flat against the surface.

2.Press Trigg while moving the pen. If calibration is performed from a menu other than the Settings main menu, you should press Trigg, hold the pen still for two seconds and then move it over the surface.

3.A progress indicator appears on C-Pen’s screen.

4.Move C-Pen over the text, in circles or back and forth. Continue to move C-Pen until the progress bar is complete.

5.A message indicates that calibration has been completed.

When your C-Pen is calibrated the Text Quality icon in the status field changes appearance.

Text Quality icon