Clarity - INT9 A/D converter

Installation

and tinned endings – red (+), white (-) and shielding (analogue ground)].

Starting (marker) cables

Labelled “IN1” to “IN4” (correlating to channel number) and tipped with CINCH connector (female).

The following cables can be connected to this CINCH connector:

A cable terminated with free leads [red (+), shielding (digital ground)] for direct connection to the chromatograph.

A cable terminated with a button for cases where a starting contact is not available and it is necessary to perform a manual start.

Both types of cable are supplied for each starting cable.

Digital output cables

Relay contacts labelled "OUT1R" to "OUT4R" (according to the number of channels), terminated with free leads. These are used for synchronizing autosamplers without an AS Control module in the Active Sequence or for controlling other devices.

At the converter end the cable is always terminated with a D-SUB 37-pin connector (female).

2.3Connection with Chromatograph

Connect the cables according to one of the following diagrams in Fig. 2. Use the symmetrical connection only when you are sure that the chromatograph/detector is equipped with a symmetrical output – it is necessary to read through the instructions for the corresponding chromatograph.

All current DataApex A/D Converters INT7, INT9, U-PAD,U-PAD2and Net-PADuse the same standard INT7 Connector.

Note: A description of the INT7 connector can be found in chapter 5.1 on pg. 32.

7

Page 7
Image 7
Clarity INT9 manual Connection with Chromatograph, Starting marker cables, Digital output cables