SPEC | Default |
|
| Bit 3 |
|
| Bit 2 |
|
|
| Bit 1 |
|
| Bit 0 |
|
|
|
|
|
|
|
|
|
|
| ||||||
27 | 0100 | Comma Display | Keypad Entry of Tare |
| Tare Range |
|
|
|
| ||||||
|
| 0: No |
|
| 0: No |
|
|
| 00: 100% FS |
| 01: | 50% FS | |||
|
| 1: Yes |
|
| 1: Yes |
|
|
| 10: 5% FS |
| 11: Not used | ||||
|
|
|
|
|
|
| |||||||||
28 | 0000 | Auto Tare Clear when | Automatic Unit Weight Clear Condition |
| Automatic Unit Weight | ||||||||||
|
| Rezeroed |
| 00: Over net 5d and gross 21d, weight stable | Clear |
|
| ||||||||
|
|
|
|
|
|
|
|
| |||||||
|
| 0: | No |
|
| 01: Greater than or equal to net 1d, weight stable | 0: No |
|
| ||||||
|
| 1: | Yes (REZERO | 10: Greater than or equal to net 1d, quantity > 0, | 1: Yes |
|
| ||||||||
|
|
| clears tare value) | weight stable |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
| ||||||||
29 | 0000 | Digital Tare Rounding | Tare Value Exchange |
| Tare Addition |
| Tare Subtraction | ||||||||
|
| 0: Tare exactly |
| with TARE key |
|
| 0:Yes (new tare weight | 0: Yes (new tare weight | |||||||
|
|
|
|
|
|
| |||||||||
|
| 1: Round to nearest | 0: Yes (allow tare |
| added to existing tare) | subtracted from | |||||||||
|
|
| increment |
| addition/subtraction) | 1:No |
|
| existing tare) | ||||||
|
|
|
|
|
| 1: No |
|
|
|
|
|
| 1: No |
|
|
|
|
|
|
|
|
|
|
|
| ||||||
30 | none | Load Cell Sensitivity Selection |
|
|
|
|
| ||||||||
|
| Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise. | |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| Spec | Min | Max | Spec | Min | Max |
| Spec | Min | Max | Spec | Min | Max | |
|
| 0000: | 3.46 | 4.00 | 0100: | 1.95 | 2.25 |
| 1000: | 1.09 | 1.27 | 1100: | 0.61 | 0.71 | |
|
| 0001: | 3.00 | 3.46 | 0101: | 1.69 | 1.95 |
| 1001: | 0.95 | 1.09 | 1101: | 0.53 | 0.91 | |
|
| 0010: | 2.59 | 3.00 | 0110: | 1.46 | 1.69 |
| 1010: | 0.82 | 0.95 | 1110: | 0.46 | 0.53 | |
|
| 0011: | 2.25 | 2.59 | 0111: | 1.27 | 1.46 |
| 1011: | 0.71 | 0.82 | ||||
|
|
| 1111: | 0.40 | 0.46 | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| ||||||||
31 | none | Load Cell Sensitivity Selection |
|
|
|
|
| ||||||||
|
| Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise. | |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| Spec | Min | Max | Spec | Min | Max |
| Spec | Min | Max | Spec | Min | Max | |
|
| 0000: | 3.46 | 4.00 | 0100: | 1.95 | 2.25 |
| 1000: | 1.09 | 1.27 | 1100: | 0.61 | 0.71 | |
|
| 0001: | 3.00 | 3.46 | 0101: | 1.69 | 1.95 |
| 1001: | 0.95 | 1.09 | 1101: | 0.53 | 0.91 | |
|
| 0010: | 2.59 | 3.00 | 0110: | 1.46 | 1.69 |
| 1010: | 0.82 | 0.95 | 1110: | 0.46 | 0.53 | |
|
| 0011: | 2.25 | 2.59 | 0111: | 1.27 | 1.46 |
| 1011: | 0.71 | 0.82 | ||||
|
|
| 1111: | 0.40 | 0.46 | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| ||||||||
32 | 1010 | Calibration Mode | Low Battery |
|
| Auto Exit from Part | Scale Connected to | ||||||||
|
| Protected by Span | Annunciator Enabled |
| Accumulation and | PORT 2 |
| ||||||||
|
| Switch |
|
| 0: Yes |
|
|
| Reduction Mode | 0: No |
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
| 0: Yes (span switch | 1: No |
|
|
| 0:No (must press | 1: Yes |
|
| |||||
|
|
| must be pressed |
|
|
|
| CLEAR to perform | CAUTION: If you are not | ||||||
|
|
| before calibration) |
|
|
|
| another accumulation) | |||||||
|
|
|
|
|
|
| connecting a scale to this | ||||||||
|
| 1: No |
|
|
|
|
|
| 1:Yes (exits to counting | ||||||
|
|
|
|
|
|
|
| connector, select 0. | |||||||
|
|
|
|
|
|
|
|
|
| mode after three | |||||
|
|
|
|
|
|
|
|
|
| Selecting 1 can damage | |||||
|
|
|
|
|
|
|
|
|
| seconds) |
| ||||
|
|
|
|
|
|
|
|
|
|
| the console’s electronic | ||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
| components. |
| |
|
|
|
|
|
|
|
| ||||||||
33 | 0___ | Over Weight Mask at: | Weight Decimal Point Position (PORT 2, Scale 2 or 3) |
|
| ||||||||||
|
| 0: +1d |
|
| MUST BE SET PRIOR TO CALIBRATION! |
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
| 1: +9d |
|
| 000: 00000 |
| 011: 00.000 |
|
|
|
|
| |||
|
|
|
|
|
| 001: 0000.0 |
| 100: 0.0000 |
|
|
|
|
| ||
|
|
|
|
|
| 010: 000.00 |
| 101: 0.00000 |
|
|
|
| |||
|
|
|
|
|
|
|
|
|
| ||||||
34 | 0000 | Not used |
| A/D Output (PORT 1, |
| A/D Filtering (PORT 1, Scale 1) |
|
| |||||||
|
|
|
|
|
| Scale 1) |
|
| 00: Normal |
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| 0: For std/normal load | 01: Protect from small vibration/fast change in | ||||||||
|
|
|
|
|
| cell | (≤ 18 mV) |
| display |
|
|
|
| ||
|
|
|
|
|
| 1: For abnormal load |
| 10: Protect from medium vibration |
|
| |||||
|
|
|
|
|
| cell with too large |
| 11: Protect from large vibration, slow change in | |||||||
|
|
|
|
|
| offset (> 18mV) |
| display |
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|