SPEC

Default

 

 

Bit 3

 

 

Bit 2

 

 

 

Bit 1

 

 

Bit 0

 

 

 

 

 

 

 

 

 

 

 

27

0100

Comma Display

Keypad Entry of Tare

 

Tare Range

 

 

 

 

 

 

0: No

 

 

0: No

 

 

 

00: 100% FS

 

01:

50% FS

 

 

1: Yes

 

 

1: Yes

 

 

 

10: 5% FS

 

11: Not used

 

 

 

 

 

 

 

28

0000

Auto Tare Clear when

Automatic Unit Weight Clear Condition

 

Automatic Unit Weight

 

 

Rezeroed

 

00: Over net 5d and gross 21d, weight stable

Clear

 

 

 

 

 

 

 

 

 

 

 

 

 

0:

No

 

 

01: Greater than or equal to net 1d, weight stable

0: No

 

 

 

 

1:

Yes (REZERO

10: Greater than or equal to net 1d, quantity > 0,

1: Yes

 

 

 

 

 

clears tare value)

weight stable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

29

0000

Digital Tare Rounding

Tare Value Exchange

 

Tare Addition

 

Tare Subtraction

 

 

0: Tare exactly

 

with TARE key

 

 

0:Yes (new tare weight

0: Yes (new tare weight

 

 

 

 

 

 

 

 

 

1: Round to nearest

0: Yes (allow tare

 

added to existing tare)

subtracted from

 

 

 

increment

 

addition/subtraction)

1:No

 

 

existing tare)

 

 

 

 

 

 

1: No

 

 

 

 

 

 

1: No

 

 

 

 

 

 

 

 

 

 

 

 

30

none

Load Cell Sensitivity Selection (mV/V)—PORT 1, Scale 1

 

 

 

 

 

 

 

Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Spec

Min

Max

Spec

Min

Max

 

Spec

Min

Max

Spec

Min

Max

 

 

0000:

3.46

4.00

0100:

1.95

2.25

 

1000:

1.09

1.27

1100:

0.61

0.71

 

 

0001:

3.00

3.46

0101:

1.69

1.95

 

1001:

0.95

1.09

1101:

0.53

0.91

 

 

0010:

2.59

3.00

0110:

1.46

1.69

 

1010:

0.82

0.95

1110:

0.46

0.53

 

 

0011:

2.25

2.59

0111:

1.27

1.46

 

1011:

0.71

0.82

 

 

 

1111:

0.40

0.46

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

31

none

Load Cell Sensitivity Selection (mV/V)—PORT 1, Scale 2

 

 

 

 

 

 

 

Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Spec

Min

Max

Spec

Min

Max

 

Spec

Min

Max

Spec

Min

Max

 

 

0000:

3.46

4.00

0100:

1.95

2.25

 

1000:

1.09

1.27

1100:

0.61

0.71

 

 

0001:

3.00

3.46

0101:

1.69

1.95

 

1001:

0.95

1.09

1101:

0.53

0.91

 

 

0010:

2.59

3.00

0110:

1.46

1.69

 

1010:

0.82

0.95

1110:

0.46

0.53

 

 

0011:

2.25

2.59

0111:

1.27

1.46

 

1011:

0.71

0.82

 

 

 

1111:

0.40

0.46

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

32

1010

Calibration Mode

Low Battery

 

 

Auto Exit from Part

Scale Connected to

 

 

Protected by Span

Annunciator Enabled

 

Accumulation and

PORT 2

 

 

 

Switch

 

 

0: Yes

 

 

 

Reduction Mode

0: No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0: Yes (span switch

1: No

 

 

 

0:No (must press

1: Yes

 

 

 

 

 

must be pressed

 

 

 

 

CLEAR to perform

CAUTION: If you are not

 

 

 

before calibration)

 

 

 

 

another accumulation)

 

 

 

 

 

 

 

connecting a scale to this

 

 

1: No

 

 

 

 

 

 

1:Yes (exits to counting

 

 

 

 

 

 

 

 

connector, select 0.

 

 

 

 

 

 

 

 

 

 

mode after three

 

 

 

 

 

 

 

 

 

 

Selecting 1 can damage

 

 

 

 

 

 

 

 

 

 

seconds)

 

 

 

 

 

 

 

 

 

 

 

 

the console’s electronic

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

components.

 

 

 

 

 

 

 

 

 

33

0___

Over Weight Mask at:

Weight Decimal Point Position (PORT 2, Scale 2 or 3)

 

 

 

 

0: +1d

 

 

MUST BE SET PRIOR TO CALIBRATION!

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1: +9d

 

 

000: 00000

 

011: 00.000

 

 

 

 

 

 

 

 

 

 

 

001: 0000.0

 

100: 0.0000

 

 

 

 

 

 

 

 

 

 

 

010: 000.00

 

101: 0.00000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

34

0000

Not used

 

A/D Output (PORT 1,

 

A/D Filtering (PORT 1, Scale 1)

 

 

 

 

 

 

 

 

Scale 1)

 

 

00: Normal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0: For std/normal load

01: Protect from small vibration/fast change in

 

 

 

 

 

 

cell

(≤ 18 mV)

 

display

 

 

 

 

 

 

 

 

 

 

1: For abnormal load

 

10: Protect from medium vibration

 

 

 

 

 

 

 

 

cell with too large

 

11: Protect from large vibration, slow change in

 

 

 

 

 

 

offset (> 18mV)

 

display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4-2. DC-190 142 Settings (Continued)

Page 17
Image 17