Digi DC-190 Load Cell Sensitivity Selection mV/V-PORT 2, Scale 2 or, Max Spec Min

Models: DC-190

1 33
Download 33 pages 24.46 Kb
Page 18
Image 18

SPEC

Default

 

 

Bit 3

 

 

 

Bit 2

 

 

 

Bit 1

 

 

Bit 0

 

 

 

 

 

 

 

 

 

 

 

35

0000

Not used

 

 

A/D Output (PORT 1,

A/D Filtering (PORT 1, Scale 2)

 

 

 

 

 

 

 

 

 

Scale 2)

 

00: Normal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0: For std/normal load

01: Protect from small vibration/fast change in

 

 

 

 

 

 

 

cell (≤ 18 mV)

 

display

 

 

 

 

 

 

 

 

 

 

 

1: For abnormal load

10: Protect from medium vibration

 

 

 

 

 

 

 

 

 

cell with too large

11: Protect from large vibration, slow change in

 

 

 

 

 

 

 

offset (> 18mV)

 

display

 

 

 

 

 

 

 

 

 

 

 

36

__00

Minimum Display (PORT 2, Scale 2 or 3)

A/D Filtering (PORT 2, Scale 2 or 3)

 

 

 

MUST BE SET PRIOR TO CALIBRATION!

00: Normal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00:

2

 

01:

1

 

 

 

01: Protect from small vibration/fast change in

 

 

10:

5

 

11:

10

 

 

 

 

display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10: Protect from medium vibration

 

 

 

 

 

 

 

 

 

 

 

 

11: Protect from large vibration, slow change in

 

 

 

 

 

 

 

 

 

 

 

display

 

 

 

 

 

 

 

 

 

 

 

 

37

1001

Load Cell Sensitivity Selection (mV/V)—PORT 2, Scale 2 or 3

 

 

 

 

 

 

Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Spec

Min

Max

Spec

Min

Max

Spec

Min

Max

Spec

Min

Max

 

 

0000:

3.46

4.00

0100:

1.95

2.25

1000:

1.09

1.27

1100:

0.61

0.71

 

 

0001:

3.00

3.46

0101:

1.69

1.95

1001:

0.95

1.09

1101:

0.53

0.91

 

 

0010:

2.59

3.00

0110:

1.46

1.69

1010:

0.82

0.95

1110:

0.46

0.53

 

 

0011:

2.25

2.59

0111:

1.27

1.46

1011:

0.71

0.82

 

 

1111:

0.40

0.46

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

38

0010

A/D Output (Scale 3)

Digital Tare with Weight

Internal Count

 

Stability Check When

 

 

0: For std/normal load

on Scale

 

0:

500,000

 

Changing Scale

 

 

 

 

 

 

 

 

 

 

 

cell (≤ 18 mV)

 

0: Yes

 

 

1:

1,000,000

 

0: Yes

 

 

 

 

1:

For abnormal load

1: No

 

 

 

 

 

 

1: No

 

 

 

 

 

cell with too large

 

 

 

 

 

 

 

 

 

 

 

 

 

offset (> 18mV)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

39

0010

Set SPEC 39 to 0010

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4-2. DC-190 142 Settings (Continued)

Page 18
Image 18
Digi DC-190 installation manual Load Cell Sensitivity Selection mV/V-PORT 2, Scale 2 or, Max Spec Min