Digi installation manual 10.11 DC-190 Ultra Count Specifications

Models: DC-190

1 33
Download 33 pages 24.46 Kb
Page 32
Image 32

10.11 DC-190 Ultra Count Specifications

Power

Power source

100–120VAC or 220–240 VAC at

 

50/60 Hz

Power consumption

20 W

Analog Specifications

Input sensitivity

0.40–4.00 mV/V

Zero adjustment range

100% FS

A/D conversion rate

10/sec

Load cell excitiation

12V DC

Load cells per scale

4 x 350Ω

Load cell connectors

2 x 14-pin Amphenol

Scale channels

Up to four: three analog (load cell)

 

scales, plus one force balance

Weightdisplay resolution

1/10000, 1/5000, 1/2500

Counting resolution

1/1000000

Digital Specifications

Weight display

5 digits

Unit weight display

5 digits

Quantity display

7 digits

Item memory

200 items, expandable with IMS

 

software

Serial Communications and Setpoint Connectors

SCANNER port

8-pin DIN, RS-232C

FORCE BALANCE port

8-pin DIN, RS-232C

232C/PRINTER port

9-pin D-Sub, RS-232C

Setpoint output

8-pin mini-DIN, 24V open collector

Environmental

Operating temperature

14°F to 104°F (–10°C ° to 40°C)

Operating humidity

10% – 80% relative humidity

Physical

Bench type

16.4 in (W) x 17.1 in (D) x 4.9 in (H)

 

41.7 cm (W) x 43.4 cm (D) x 12.4 cm (H)

 

17.6 lb (8.0 Kg)

Pole type

16.4 in (W) x 16.5 in (D) x 17.9 in (H)

 

41.7 cm (W) x 41.9 cm (D) x 45.5 cm (H)

 

19.4 lb (8.8 Kg)

Page 32
Image 32
Digi installation manual 10.11 DC-190 Ultra Count Specifications