Dolby Laboratories CP65 A-ChainAlignment Procedures, b. Optical Preamplifier Adjustments, Action

Models: CP65

1 82
Download 82 pages 36.23 Kb
Page 31
Image 31
A-Chain Alignment Procedures

 

4-8

 

 

 

A-Chain Alignment Procedures

 

 

 

b. Optical Preamplifier Adjustments

 

 

 

 

 

 

Step

Action

Indication

 

No.

 

 

 

 

4

Cat. No.

 

 

240A

 

 

Proj. 1

Manual background GAIN Lt

hf

Manual background GAIN

Rt

hf

Lt tp

Lt

signal

present

Rt

Rt tp

Manual background GAIN Lt

hf

Manual background GAIN Rt

hf

Proj. 2 gnd

5

format

01Mono

format

01Mono

6

 

 

 

RTA

Cat. No. 240A

 

 

 

 

 

IN

Proj. 1

Status

 

 

RV101

Gain

L

 

RV102 hf

 

 

 

RV201

Gain

R

 

RV202 hf

 

 

 

TP501 L tp

 

X

 

L

SIGNAL

 

 

 

 

 

 

PRESENT

Y

 

R

 

 

 

 

TP502 R tp

 

 

RV301 Gain

L

 

RV302 hf

R

 

RV401 Gain

SCOPE

RV402 hf

 

Proj. 2

Status

 

DUAL TRACE

TP503 GND

 

 

MODE

 

 

 

NOTE: Be sure that the vertical range is set the same on both channels.

4-9

A-Chain Alignment Procedures

b. Optical Preamplifier Adjustments

Notes

On the Cat. No. 240A Optical Preamplifier Card

Turn all of the Proj 1 and Proj 2 hf potentiometers fully counterclockwise (CCW).

Verify that power to the power amplifiers is OFF.

Press the Mono 01 format switch on the front panel; the LED in the switch should light.

Connect the RTA and the oscilloscope to the Cat. No. 240A TP501 L (left) and TP502 R (right) test points as shown in the interconnection diagram. Switch the scope to dual trace mode.

Earth (ground) both instruments at TP503 GND on the card.

The input to the RTA will be switched to the left channel or right channel in the following steps.

Page 31
Image 31
Dolby Laboratories CP65 manual A-ChainAlignment Procedures, b. Optical Preamplifier Adjustments, Action, Indication