Emerson 00825-0100-4530 Rev EC manual Canadian Standards Association CSA approvals

Models: 00825-0100-4530 Rev EC

1 36
Download 36 pages 54.13 Kb
Page 26
Image 26

Quick Installation Guide

March 2013

 

 

I5, IE1 Intrinsically safe and FISCO model:

Intrinsically Safe for Class I, II, III, Div. 1, Groups A, B, C, D, E, F, and G,

Class I, Zone 0, AEx ia IIC T4 when installed per Control Drawing: 9240 030-936. Non-Incendive Class I, Div. 2, Groups A, B, C, D, F, and G;

Suitable for Class II, III, Div. 2, Groups F and G;

4-20 mA/HART model: Ui = 30 Vdc, Ii = 130 mA, Pi = 1.0 W, Ci = 7.26 nF, Li = 0 H.

Max. operation: 42.4 V, 25 mA

FOUNDATION fieldbus model: Ui = 30 Vdc, Ii = 300 mA, Pi = 1.3 W, Ci = 0 nF, Li = 0 H.

Max. operation: 32 V, 25 mA

FISCO model: Ui = 17.5 Vdc, Ii = 380 mA, Pi = 5.32 W, Li = Ci = 0. Temp. Code T4

Ambient temperature limits: -50 °C to +70 °C2

Approval valid for HART, FOUNDATION fieldbus, and FISCO options.

Canadian Standards Association (CSA) approvals I5, IE1 Intrinsically safe and FISCO model:

Certificate: 1514653

Product options bearing the Dual Seal marking meets the Dual Seal Requirements of ANSI/ISA12.27.01-2003.

Dual seal annunciation

The breach of the secondary seal is annunciated via product leakage from the antenna’s vents. The leakage will be visible and/or audible from the transmitter

head’s threads.

Dual seal maintenance

No maintenance required. Verify proper operation by keeping leak path free from ice or contamination.

WARNING: The substitution of components may impair intrinsic safety.

E61 Explosion-proof with internal intrinsically safe circuits [Exia]

Class I, Div. 1, Groups B, C, and D;

Class II, Div. 1 and 2, Groups E, F, and G;

Class III, Div. 1

Temp Code T4.

Ambient temperature limits -50 °C to +70 °C2

Approval valid for HART, FOUNDATION fieldbus, and Modbus options.

I6, IF1 Intrinsically Safe Exia:

Class I, Div. 1, Groups A, B, C, and D. Temp Code T4.

4-20 mA/HART model: Ui = 30 Vdc, Ii = 130 mA, Pi = 1.0 W, Ci = 7.26 nF, Li = 0 H.

FOUNDATION fieldbus model: Ui = 30 Vdc, Ii = 300 mA, Pi = 1.3 W, Ci = 0 nF, Li = 0 H.

FISCO model: Ui = 17.5 Vdc, Ii = 380 mA, Pi = 5.32 W, Li = Ci = 0.

Installation Drawing: 9240 030-937 Ambient temperature limits -50 °C to +70 °C2.

Approval valid for HART, FOUNDATION fieldbus, and FISCO options.

1)Ordering information code for Product Certifications, see the Rosemount 5300 Series Product Data Sheet (Document No. 00813-0100-4530) or the Rosemount 5300 Series Reference Manual

(Document No. 00809-0100-4530).

2)+60 °C with FOUNDATION fieldbus or FISCO option.