SECTION 2

Instruction Bulletin

World Class 3000

106-300NE Rev. 3.4 May 2000

SECTION 2

INSTALLATION

2

2-1 OXYGEN ANALYZER (PROBE)

either make necessary repairs or install

INSTALLATION

the probe upstream of any leakage.

Before starting to install this equip- ment, read the "Safety instructions for the wiring and installation of this ap- paratus" at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in serious injury or death.

Install all protective equipment covers and safety ground leads after installa- tion. Failure to install covers and ground leads could result in serious injury or death.

a.Selecting Location

1.The location of the probe in the stack or flue is most important for maximum accuracy in the oxygen analyzing pro- cess. The probe must be positioned so that the gas it measures is representa- tive of the process. Best results are normally obtained if the probe is posi- tioned near the center of the duct (40 to 60% insertion). A point too near the edge or wall of the duct may not pro- vide a representative sample because of the possibility of gas stratification. In addition, the sensing point should be selected so that the process gas tem- perature falls within a range of 50° to 1300°F (10° to 704°C). Figure 2-1pro- vides mechanical installation references.

2.Check the flue or stack for holes and air leakage. The presence of this con- dition will substantially affect the accu- racy of the oxygen reading. Therefore,

3.Ensure that the area is clear of ob- structions internal and external that will interfere with installation. Allow ade- quate clearance for removal of probe (Figure 2-1).

Do not allow the temperature of the probe junction box to exceed 300°F (149°C) or damage to the unit may re- sult. If the probe junction box tempera- ture exceeds 300°F (149°C), the user must fabricate a heat shield or provide adequate cooling air to the probe junc- tion box.

b.Mechanical Installation

1.Ensure that all components are avail- able for installation of the probe. Ensure that the system cable is the re- quired length. If applicable, check the ceramic filter to ensure that it is not damaged.

2.The probe may be installed intact as it is received. It is recommended that you disassemble the adapter plate for each installation.

NOTE

An abrasive shield is recommended for high velocity particulate in the flue stream (such as those in pulverized coal kilns and recovery boilers). Verti- cal and horizontal brace clamps are provided for 9 ft and 12 ft (2.75 m and

3.66m) probes to provide mechanical support of the probe. Refer to Figure 2-1, Sheet 5.

3.Weld or bolt adapter plate (Figure 2-1)onto the duct.

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-1