Figure A-16. Cell Wiring Connection

World Class 3000

Instruction Manual

Appendix A Rev. 3.6

 

July 1998

A

 

INCONEL

HEATER

 

CELL WIRE

WIRES

 

(CLEAR

(BLACK

THERMOCOUPLE -

SLEEVING)

SLEEVING)

(RED ALUMEL)

BOMB TAIL

CONNECTOR Manual background

CELL EXTENSION Manual background

WIRE (ORANGE)

THERMOCOUPLE + (YELLOW CHROMEL)

CALIBRATION GAS FITTING

REFERENCE AIR FITTING

CABLE

PROBE JUNCTION

BOX COVER

27270019

Figure A-16. Cell Wiring Connection

b.If the probe uses the standard diffusion element, use a spanner wrench to remove the diffusion element.

c.If equipped with the optional ceramic diffusor assembly, remove and discard setscrews, Figure A-17,and remove vee deflector. Use spanner wrenches from probe disassembly kit, Table A-3,to turn hub free from retainer. Inspect diffusion element. If damaged, re- place element.

d.Loosen four socket head cap screws from the cell and flange assembly and remove the assembly and the corrugated seal. The cell flange has a notch which may be used

to gently pry the flange away from the probe. Note that the contact pad inside the probe will sometimes fuse to the oxygen sensing cell. If the cell is fused to the con- tact pad, push the cell assembly back into the probe (against spring pressure), and quickly twist the cell assembly. The cell and contact pad should separate. If the contact pad stays fused to the cell, a new con- tact/thermocouple assembly must be in- stalled. Disconnect the cell and the thermocouple wires at the probe junction box, and withdraw the cell with the wires still attached (see paragraph A-13).