Emerson manual Rosemount 3051S, Reference Manual, AI Block FFAIRMT3, Auto, 3-19

Models: 3051S

1 32
Download 32 pages 28.03 Kb
Page 19
Image 19
AI Block

Reference Manual

00809-0200-4801, Rev BA April 2007

Rosemount 3051S

Display Parameter 1

Block Type #1

Block Tag #1

AI Block

FFAI_RMT3

Param Index #1

OUT

Custome Tag #1

Units Type #1

Custom Units #1

Auto

DISPLAY_PARAM_SEL

The DISPLAY_PARAM_SEL parameter specifies how many process variables will be displayed. Select up to four display parameters.

BLK_TAG_#(1)

Enter the Block Tag of the function block that contains the parameter to be displayed.

BLK_TYPE_#(1)

Enter the Block Type of the function block that contains the parameter to be displayed. This parameter is generally selected via a drop-down menu with a list of possible function block types. (e.g. Transducer, PID, AI, etc.)

PARAM_INDEX_#(1)

The PARAM_INDEX_# parameter is generally selected via a drop-down menu with a list of possible parameter names based upon what is available in the function block type selected. Choose the parameter to be displayed.

CUSTOM_TAG_#(1)

The CUSTOM_TAG_# is an optional user-specified tag identifier that can be configured to be displayed with the parameter in place of the block tag. Enter a tag of up to five characters.

UNITS_TYPE_#(1)

The UNITS_TYPE_# parameter is generally selected via a drop-down menu with three options: AUTO, CUSTOM, or NONE. Select AUTO only when the parameter to be displayed is pressure, temperature, or percent. For other parameters, select CUSTOM and be sure to configure the CUSTOM_UNITS_# parameter. Select NONE if the parameter is to be displayed without associated units.

(1)_# represents the specified parameter number.