1.2System Specifications Table 1-1 lists the BioOptix 10 specifications.

 

Table 1-1 BioOptix 10 Specifications

 

 

 

Dimensions a

 

Height: 69 cm

 

 

Width: 72 cm

 

 

Depth: 61 cm

 

 

 

Weight a,b

 

210 kg

Power Requirement a,c

 

100 VAC ±10%, 2.0 Amps

The power switch is the

 

117 VAC ±10%, 2.0 Amps

 

234 VAC ±10%, 1.0 Amps

disconnect device.

 

 

 

 

 

 

Line Frequency

 

50 or 60 Hz

 

 

 

Operating Temperature a

 

5 to 40°C

Humidity a

 

When connected to mains power, 95% relative humidity maximum at 5 to

 

 

40°C

 

 

 

Gradient Flow Rate Range

 

0.5 to 12 mL/min

 

 

 

Flow Rate Accuracy

 

± 5% full scale (using water)

 

 

 

Gradient Accuracy

 

± 5% full scale (using water)

 

 

 

Gradient Linearity

 

± 5% full scale (using water)

 

 

 

Operating Pressure

 

0 to 6.895 bar (0 to 100 psi)

 

 

 

Flow Cell Path Length

 

2 mm

 

 

 

Detector Wavelength

 

190 to 360 nm

 

 

 

Detector Noise

 

5 10-4AU

Detector Drift

 

1 10-3AU/hr

CE Conformity Specifications a,d

 

Pollution Degree: 2

 

 

Installation Category: II

 

 

Maximum Altitude: 2000 meters

 

 

 

a.Excludes Foxy 200 Fraction Collector and Windows PC Controller. Refer to the respective manuals for these additional specifications.

b.Dry weight of the BioOptix 10, excluding connection cables, tubing, and user-supplied buffer containers.

c.Operating voltage of the BioOptix 10 is factory set.

d.Refer to the CE Declaration of Conformity, at the back of this manual, for applicable standards and test results.