BioOptix™ 10

Section 2 Installation

Common Buffer Inlet

Mounting

Block

Manifold

Figure 2-14 Buffer assembly routes a common solution to all ten channels

Figure 2-15 Buffer assemblies installed

To connect the Buffer Assembly:

a.Remove the two screws located in the center of the Buffer A inlet ports. (The two exposed holes are used to mount the Buffer Assembly.)

b.Attach the mounting block to the two holes using the hex head screws.

c.Attach the manifold to the mounting block with the Phillips screws and lock washers.

d.Connect the ten Teflon tubes (each 3" long by 0.085 ID) from the manifold to the Buffer A inlet ports.

e.Cut to length a section of 0.250" OD Teflon tubing (P/N 023-0508-01) from the accessory kit and route the buffer solution to the common inlet. The standard length is 60" or 152 cm.

2-11

Page 29
Image 29
Teledyne 10 manual Common Buffer Inlet Mounting Block Manifold