BioOptix™ 10
Section 2 Installation
2-16
6. Ensure that each channel on the BioOptix 10 has
0.125" OD bypass tubing between the Column Inlet and
Outlet ports (Figure 2-18).
Figure 2-18 Column bypass tubing
7. For each channel, set the mode control knobs to Purge (Fig-
ure 2-19).
Figure 2-19 Mode control knob
8. From the BioOptix 10 software menu, select Tools>Manual
Control. The BioOptix 10 Manual Hardware Control win-
dow is displayed (Figure 2-20).
Column Inlet
Column Outlet