Section 2 Installation
10.Connect the tubing from the column module to the pump module. Refer to Figure
a.Remove the two thumbscrews securing the splash guard to the pump module. Set aside the splash guard.
Note
The tubing numbers from the column module match the num- bers above the compression fittings on the pump module. (The tubing is visible in Figure
b.Loosen the nut on the compression fitting, insert the tubing fully into the fitting and
c.Replace the splash guard so that the tubing is routed up and over the guard.
Fittings for column module tubing
Splash Guard
Thumbscrew
Figure 2-10 Liquid connections (pump module shown)
11.Slide the two modules together (Refer to Figure
Note
Use care to avoid damaging internal tubing and cables. Do not pinch or crimp tubing or cables between modules.