BioOptix™ 10 Appendix A Method Templates
A.2 Modifying the
Template
5.Check the box labeled Always use this program to open these files.
6.Click OK.
After doing so,
Once open in a text editor, the template file can be modified. Refer to Figure
[Blocks]
Count=5
[Block1]
Name="Equilibrate"
Type="Fluid"
AskPause="Yes"
.
.
.
[Other]
Wavelength=280
;Use 254 nm for
;nucleic acids.
Block Count = total number of blocks (value range: 1 to 10)
Name = Descriptive block title (length: 0 to 20 characters)
Type = Block function (options: Fluid, Load, Gradient)
AskPause = Pause after completion? (options: Yes, No)
Wavelength = Detection wavelength (value range: 190 to 360)
Comment = precede notes with a semicolon as the first character on each line.
Figure A-2 Customizable values in the Standard.tpl file
| After modifying the template, save the file. If saved as |
| Standard.tpl in the same folder as the biooptix.exe program file, |
| all new method files will be created according to your changes. |
A.2.1 General Editing Rules | There are three types of changes that can be made to the file. |
| • Numeric values – Numeric values follow an equal sign |
| and do not use quotes. Enter a numeric value within the |
| acceptable range. |
| Count=5 in Figure |
| • Text entries – Text entries are enclosed in quotes and |
| follow an equal sign. Enter text between the quotes, |
| observing the string length or option requirements. Any |
| printing ASCII character may be used. Nonprinting |
| characters such as returns, line feeds, tabs, must be |
| avoided. Entries are not |
| Name=”Equilibrate” in Figure |
| • Comments – Comments can be added to the file to |
| annotate changes, special instructions, author info, etc. |
| ; Use 254 nm… in Figure |