BioOptix™ 10

Section 2 Installation

To keep tubing length at a minimum, buffer, sample, and waste collection containers should be placed to the left of the unit. The containers may also be kept behind the unit as long as the con- tainers are accessible.

The Foxy 200 should be installed to the right of the BioOptix 10. The fraction collector requires about 1800 square centimeters (275 in2) of bench space. Ensure that the full motion of the fraction collector arm is not obstructed.

The Windows PC may be placed anywhere within 2 meters of the

BioOptix 10 unit (the length of the serial control cable).

=Keep area clear for instrument connections, ventilation and to access On/Off Switch.

100 cm

16 cm

8 cm

(Rear)

33 cm

50 cm

16cm

8 cm

56 cm

 

BioOptix

61 cm

 

13 cm

Foxy 200

 

 

 

10 cm

(Front)

 

10 cm

44 cm

12 cm

29 cm

 

 

 

Figure 2-1 System space requirements

2.2.2

Electrical

The operating voltage of the BioOptix 10 is factory set. Refer to

 

Requirements

the serial number label on the right side panel to determine the

 

 

electrical requirements.

 

 

Teledyne Isco recommends that the BioOptix 10 system runs on

 

 

a dedicated AC circuit to prevent abnormal conditions that may

 

 

be caused by high current or electrical noise. The BioOptix 10

 

 

and Foxy 200 should be connected to the same circuit to prevent

 

 

ground loop anomalies.

2.2.3

Assembling the

The BioOptix 10 is shipped in two pieces—a column module and

 

BioOptix 10

a pump module. After unpacking and inspecting the modules

 

 

place them as shown in Figure 2-2 on a sturdy work surface. This

 

 

will allow access to the internal electrical and fluid connections.

2-2

Page 20
Image 20
Teledyne 10 manual Electrical, Assembling, BioOptix