BioOptix™ 10 Section 4 System Maintenance
4.4Deuterium Lamp Replacement
17.Hold the push rod
18.Refer to Figure
19.Repeat steps 10 through 18 for the remaining channels.
Note
Channel six is located below the inductor. To remove the glass cylinder on this channel, first remove the inductor by unfasten- ing the #8 pan head screw and internal lock washer.
20.After replacing the seals and
a.Insert and tighten the pan head screw with lock washer into the top mounting tab (Figure
b.Insert and tighten the Allen head screw with lock washer into the bottom mounting tab.
c.Insert and tighten the four screws with lock washer through the side of the case (item 4, Figure
21.Place the drip tray (item 3, Figure
22.Return the upper valve assembly (item 2, Figure
23.Place the back cover (item 1, Figure
Note
The seal replacement procedure is complete. Perform a pump flush (see section 4.1) to prime the system.
If peak amplitude of a standard sample (see section 4.5, Trouble- shooting) degrades over time, or if the baseline is noisy on all channels, the deuterium lamp may need to be replaced.
WARNING
Risk of electric shock. Disconnect power before servicing. Only trained personnel may open the case enclosure.
WARNING
Risk of damage to eyesight and skin. Wear ultraviolet (UV) light eye protection and protect skin from exposure if the deuterium lamp is operating.