Procedure 6. A6 Power Supply Assembly

15.Disconnect all cables from the A6 power supply assembly. See Figure 4-11.

16.Use a TORX screwdriver to remove the hardware from the shield wall, the heatsink, and the A6 power supply assembly.

17.Remove the A6 power supply assembly by lifting from the regulator heatsink toward front of spectrum analyzer.

Replacement

1.Ensure that the bottom shield wall is in place before replacing the A6 power supply assembly.

2.Attach the A6 power supply assembly to the spectrum analyzer chassis and top shield wall using the four screws.

3.Connect Wl to A6J1, W3 to A6J2, fan power wires to A6J3, W8 to A6J4, and the line-power jack to A6JlOl. See Figure 4-11.

4.Secure the A6Al high voltage assembly to the A6 power supply assembly, using three TORX screws. Connect ribbon cable A6AlWl to A6J5.

5..Snap post-accelerator cable A6AlW3 to the CRT assembly.

4-22 Assembly Replacement