404 AMPL 10K Unable to adjust amplitude of 10 kHz resolution bandwidth.

Errors 405 to 416: When these 10K resolution bandwidth (RBW) error messages appear, use the following steps to check for errors 581 or 582.

1.Press [LINE to turn spectrum analyzer off.

2.Press ILINE) to turn spectrum analyzer on and observe the lower right-hand corner of the display for 10 seconds.

3.If ERR 581 or ERR 582 appears, the fault is most likely caused by the cal oscillator. Refer to errors 581 and 582.

4.If ERR 581 or ERR 582 does not appear, troubleshoot the A5 IF assembly.

Multiple IF errors during IF adjust: If a FULL IF ADJ sequence (pressing (CAL) and F0LL IF ADJ ) results in IF errors while displaying IF ADJUST STATUS: AMPLITUDE, the cal oscillator on the A4 assembly might not be providing the correct output signal. Perform the following steps:

1.Disconnect W30 (white) from A5J4.

2.Connect W30 to the input of a second spectrum analyzer and set its controls as follows:

 

 

 

Center frequency

10.7 MHz

 

 

 

Reference level

. -30 dBm

 

 

 

3. Observe the spectrum analyzer display while pressing FULL IF ADJ

 

 

 

on the spectrum analyzer. If a -35 dBm signal does not appear,

 

 

 

troubleshoot the cal oscillator on the A4 assembly.

 

 

 

 

4. If a -35 dBm signal does appear, troubleshoot the A5 IF assembly.

405

RBW

10K

Unable to adjust 10 kHz resolution bandwidth in first crystal pole.

406

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in second crystal pole.

407

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in third crystal pole.

408

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in fourth crystal pole.

409

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in first crystal pole.

410

RBW

10K

Unable to adjust 10 kHz resolution bandwidth in second crystal pole.

411

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in third crystal pole.

412

RBW

1OK

Unable to adjust 10 kHz resolution bandwidth in fourth crystal pole.

413

RBW

IOK

Unable to adjust 10 kHz resolution bandwidth in first crystal pole.

414

RBW

10K

Unable to adjust 10 kHz resolution bandwidth in second crystal pole.

General Troubleshooting 7-29