Using Frequency Response Adjustment Software

Loading the Program

Load HP BASIC into the computer. HP BASIC choices are:

nBASIC 2.0 and extensions 2.1

w BASIC 3.0 or 4.0, which must include the following binaries:

MAT

IO

GRAPH

GRAPHX

PDEV

HPIB

MS

CLOCK

CS80

DISC

KBD

For configuration instructions, refer to the BASIC Operating Manual. Next, insert the frequency response adjustment software disk into the disk drive, then type:

LOAD "FReEND8562" ,1

Note

A double-sided disk drive must be used. The frequency response adjustment

 

software will not fit on a single-sided formatted disk.

 

 

Press (EXECUTE) on HP 9000 Series 200 computers, or (i%i%i) on HP 9000 Series 300 computers, to load the software and start the program running.

Program Operation

The frequency response adjustment consists of three menus. They are the Conditions Menu, the Adjust Menu, and the Sensor Utilities Menu, which are accessed from the Conditions Menu. Program operation is controlled through a combination of softkeys and user prompts. Some prompts, primarily in the Conditions and Sensor Utilities Menus, require computer keyboard entries. Terminate keyboard entries with the (-1 or (ENTER) key. Most prompts, however, tell the user what to do next or provide informational messages.

If the message (any key) follows a prompt, pressing any key on the keyboard continues the program. If the message (any key or ‘9’ to quit) follows a prompt, pressing any key except 9 continues the program. Pressing Q terminates the current procedure at the next, most logical point in the program.

Frequency Response Adjustment Software 3-5