Agilent Technologies 856290216 service manual Al3 Second Converter

Models: 856290216

1 482
Download 482 pages 3.94 Kb
Page 161
Image 161

Al3 Second Converter

Caution Turn off the spectrum analyzer power when replacing the Al3 second converter assembly. Failure to turn off the power may result in damage to the assembly.

Removal

1.Place the spectrum analyzer upside-down on the work bench.

2.Disconnect W33, coax 81, and W35, coax 92, from the Al3 assembly.

3.Disconnect W48, coax 8, from A13J3.

4.Disconnect W57 from A13Jl.

5.Remove the four screws securing Al3 to the main deck and remove the assembly.

6.Disconnect ribbon cable W13 from the Al3 assembly.

Replacement

1.Connect ribbon cable W13 to the Al3 assembly.

2.Secure Al3 to the spectrum analyzer main deck, using four panhead screws.

3.Connect W33, coax 81, to A13J4 600 MHz IN jack.

4.Connect W35, coax 92, to A13J2 310.7 MHz OUT jack.

5.Connect W48, coax 8, to A13J3. Route W48 under W35, coax 92.

6.Connect W57 to A13Jl. Ensure that all of the connections on W56/FL2/W57 and on W48 are torqued to 113 Ncm (10 in-lb).

4-36 Assembly Replacement