Procedure 8. A7 through Al3 Assemblies

A separate replacement procedure is supplied for each assembly listed below. Before beginning a procedure, do the following:

nFold out the Al4 and Al5 assemblies as described in “Procedure 9. Al4 and Al5 Assemblies.”

nIf the All YTO or A10 assembly (SYTF or RYTHM) is being removed, also fold down the A2, A3, A4, and A5 assemblies as described in “Procedure 5. A2, A3, A4, and A5 Assemblies.”

A7 First LO Distribution Amplifier

A8 Low Band Mixer

A9 Input Attenuator

A10 YIG-Tuned Filter/Mixer (RYTHM)

All YTO

Al3 Second Converter

Figure 4-14 illustrates the location of assemblies and major cable locations.

Note

Use a torque wrench to tighten all SMA connectors to 113 Ncm (10 in-lb).

 

HP part number 8710-1655 can be used for this purpose. The style of torque

 

wrench may vary, but in all cases do not tighten the connectors beyond the

 

point at which the torque wrench “clicks” or “breaks-away.”

 

 

4-26 Assembly Replacement