6. Sampling Oscillator Adjustment

Assembly Adjusted

Al5 RF assembly

Related Performance Test

There is no related performance test for this adjustment procedure.

Description

The sampling oscillator tank circuit is adjusted for a tuning voltage of 5.05 Vdc when the sampling oscillator is set to 297.222 MHz. The voltage monitored is actually the tuning voltage divided by 4.05. The setting is then checked at other frequencies for the full tuning range of the sampling oscillator.

 

SPECTRUM ANALYZER

DIGITAL VOLTMETER

- PC

BOARD

 

 

PROP

S K 1 7

Figure 2-11. Sampler Adjustment Setup

Equipment

Digital voltmeter . . . . . . . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A DVMtestleads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 34118A

Procedure

1.Press @ to turn the spectrum analyzer off and disconnect the line power cord. Remove the spectrum analyzer cover and fold down the Al5 RF and Al4 frequency control assemblies. Prop up the Al4 frequency control assembly. Reconnect the line power cord and press (LINE] to turn the spectrum analyzer on. Connect the equipment as illustrated in Figure 2-11.