Table 1-4. Recommended Test Equipment (continued)

 

Instrument

 

Critical Specifications

Recommended

Use

 

 

 

for Equipment Substitution

Model

 

 

Xher Equipment

 

 

 

 

 

 

2ontroller

Required to run operation verification software.

HP 9816A,

V

 

 

No substitute.

HP 9836A/C,

 

 

 

 

 

HP 310,320

 

 

 

 

 

HP 332,360

 

 

1scilloscope

Bandwidth (3 dB): dc to 100 MHz

HP 54501A*

PAT

 

 

Two channels

 

 

 

 

 

Minimum vertical deflection factor: 55 mV/div

 

 

 

 

 

Minimum timebase setting: <lOO ns

 

 

 

 

 

Digitizing display with time cursors

 

 

 

 

 

Delta-t cursor accuracy in 500 ns/Div: <O.l ps

 

 

 

 

Lmplifier

Frequency range: 2.0 to 8.0 GHz

HP

11975A

P

 

 

Minimum output power (leveled)

 

 

 

 

 

2.0 to 8.0 GHz: +16 dBm

 

 

 

 

 

Output

SWR (leveled): <1.7

 

 

 

 

‘ower supply

Output

voltage: 224 Vdc

HP 6114A

A

 

 

Output voltage accuracy: <f0.2 V

 

 

 

 

ignature multimeter

Clock frequency >lO MHz

HP 5005A/B

T

 

 

Time interval function

 

 

 

 

)igital voltmeter

Range: -15 Vdc to +120 Vdc

HP 3456A*

A,T

 

 

Accuracy: <fl mV on 10 V range

 

 

 

 

 

Input impedance: 21 M D

 

 

 

 

?robes

 

 

 

 

 

 

)VM test leads

236 inches, alligator clips, probe tips

HP

34118A

A,T

 

ligh frequency probe

No substitute

HP

85024A

T

 

 

 

 

 

 

 

Part of microwave workstation

= performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting;

7 = operation verification

1-12 General Information

Page 35