4. IF Amplitude Adjustments

Equipment

Frequency synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3335A

Adapters

Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1476

Cables

BNC,122cm(48in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A Test cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85680-60093

A 5 I F

A 4 L O G A M P / C A L O S C

J9

J 5

J8 (REVIS ION CONNEC TOR

J7 J 6

J4

J 3

c707

R 4 4 5

JIO

(REVISION

CONNECTOR)

Jll

J 5

Jl

fd’J4

R53l-

sjlle

 

Figure 2-8. IF Amplitude Adjustment Locations

 

 

Note

The 15 dB reference attenuator adjustment is preset at the factory and need

 

not be done if the entire A5 IF assembly is replaced.

 

 

Procedure

1.Press luNE) to turn the spectrum analyzer off. Remove the spectrum analyzer cover and place the spectrum analyzer in the service position as illustrated in Figure 2-7.

2.Disconnect W29, violet coax cable, from A5J3. Connect the test cable between A5J3 and the 50 R output of the HP 3335A. Press m to turn the spectrum analyzer on.

Adjustment Procedures 2-27