202 FADC CAL Binary search failed during FADC linear offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

203 FADC CAL Binary search failed during FADC log offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

204 FADC CAL Binary search failed during FADC log expand offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

205 FADC CAL Slope derivation failed during FADC linear offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

206 FADC CAL Slope derivation failed during FADC log offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

207 FADC CAL Slope derivation failed during FADC log expand offset calibration.

This error applies only to spectrum analyzers with fast ADC (Option 007).

250 OUTOF RG ADC input is outside of ADC range.

251 NO IRQ Microprocessor not receiving interrupt from ADC.

LO and RF Hardware/Firmware Failures (300 to 399)

YTO Loop Errors (300 to 301)

These errors often require troubleshooting the Al4 frequency control assembly (synthesizer section) or the ADC circuits.

300 YTO UNLK YTO (first LO) phase locked loop is unlocked. The ADC measures YTO-ERR voltage under phase-lock condition.

301 YTO UNLK YTO (first LO) phase locked loop is unlocked. Same as ERR 300 except ERR 301 is set if the voltage is outside certain limits.

Roller PLL Errors (302 to 316)

These errors indicate a faulty roller oscillator on the Al4 frequency control assembly. Refer to Chapter 11, “Synthesizer Section.” The A3 interface ADC circuits may also be faulty. If error codes 333 and 499 are present, suspect the 10 MHz reference, the A21 OCXO, or on the Al5 assembly (Option 103). These errors do not apply to the hardware in an HP 8560 E-Series spectrum analyzer. If they occur in an HP 8560 E-Series spectrum analyzer, suspect a problem with the model number identification in the spectrum analyzer firmware.