2. Display Adjustment

A2R215

A2R263

AZR262

AZR268

DGTLY GAlN

START BLANK

STOP BLANK

VIDEO GAIN

AZR206

DGTLY GAIN

SK14

Figure 2-4. A2 Display Adjustment Locations

Fast Zero Span Adjustments

22.Set A2R209 SWEEP OFFSET, A2R218 VIDEO OFFSET, and A2R268 SWEEP GAIN to midrange. Adjustment locations are shown in Figure 2-4 for these A2 adjustments.

23.Set the HP 355D to 30 dB attenuation.

24.Press (PRESET) on the spectrum analyzer, and connect the equipment as shown in Figure 2-2. Set the spectrum analyzer controls as follows:

Centerfrequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..300MHz Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OHz Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40 dBm Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz Videobandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300Hz Sweep time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50ms

Adjustment Procedures 2-19

Page 59
Image 59
Agilent Technologies 856290216 service manual Fast Zero Span Adjustments, A2 Display Adjustment Locations