120 NOP ONOF ON/OFF are not valid arguments for this command.

121 NOP ARG AUTO/MAN are not valid arguments for this command.

122 NOP TRC Trace registers are not valid for this command.

123 NOP ABLK A-block format not valid here.

124 NOP IBLK I-block format not valid here.

125 NOP STRNG Strings are not valid for this command.

126 NO ? This command cannot be queried.

127 BAD DTMD Not a valid peak detector mode.

128 PK WHAT? Not a valid peak search parameter.

129 PRE TERM Premature A-block termination.

130 BAD TDF Arguments are only for TDF command.

131 ?? AM/FM AM/FM are not valid arguments for this command.

132 !FAV/RMP FAV/RAMP are not valid arguments for this command.

133 !INT/EXT INT/EXT are not valid arguments for this command.

134 ??? ZERO ZERO is not a valid argument for this command.

135 ??? CURR CURR is not a valid argument for this command.

136 ??? FULL FULL is not a valid argument for this command.

137 ??? LAST LAST is not a valid argument for this command.

138 !GRT/DSP GRT/DSP are not valid arguments for this command.

139 PLOTONLY Argument can only be used with PLOT command.

140 ?? PWRON PWRON is not a valid argument for this command.

141 BAD ARG Argument can only be used with FDIAG command.

142 BAD ARG Query expected for FDIAG command.

143 NO PRESL No preselector hardware to use command with.

144 COUPL?? Invalid COUPLING argument, expected AC or DC.

ADC Errors (200 to 299)

200SYSTEM

201SYSTEM

These errors are directly related to the ADC/interface section. Suspect a faulty A2 controller, A3 interface assembly, or Al6 fast ADC (FADC) assembly (Option 007).

Errors 202 through 207 apply only to spectrum analyzers with fast ADC (Option 007).

ADC driver/ADC hardware/firmware interaction; check for other errors.

ADC controller/ADC hardware/firmware interaction; check for other errors.