Test Equipment

Table 3-l summarizes the equipment required to run the Frequency Response Adjustment procedure. The adjustment can use various model numbers of a particular equipment type. Information about selecting the equipment model number you want to use is provided in “Setting HP-IB Addresses” in this chapter under “Frequency Response Adjustment Software.”

Note

The validity of the Frequency Response Adjustment depends in part on

 

required test equipment measurement accuracy. Verify proper calibration of

 

test equipment before adjusting the analyzer with the software.

 

 

Warmup Time

Test Equipment Warmup

Allow sufficient warmup time for test equipment. Refer to their individual operating and service manuals for warmup specifications.

Spectrum Analyzer Warmup

Warm the spectrum analyzer up for at least fifteen minutes before performing the first adjustment.

Table 3-1. Required Test Equipment Summary

Type of Equipment

Controller*

HP Model Number

HP 9000 model 216 (HP 9816), or HP 9000 model 236 (HP 9836), or HP 9000 model 310

Synthesized sweeper (10 MHz to 13.2 GHz)

HP 8340A/B, or HP 83640A, or HP 83650A

Measuring receiver

HP 8902A

Power meter

HP 436A or HP 438A (alternate)

Power sensor (10 MHz to 18 GHz)

HP 8481A. or HP 8485A

Power splitter (dc to 18 GHz)

HP 11667A, or HP 11667B

Miscellaneous cables and adapters

As per test setup

* 1 Megabyte of free memory is required for the test program.