5. DC Log Amplifier Adjustments

There are three DC log adjustments; limiter phase, linear fidelity, and log fidelity.

Assembly Adjusted

A4 log amp/Cal oscillator

Related Performance Tests

IF Gain Uncertainty

Scale Fidelity

Description

These three adjustment need only be done under the following conditions:

Limiter

phase

Only if a repair is made to blocks F, G, H, I, or J.

Linear

fidelity

Only

if

a

repair

is

made

to blocks C, D, F, G, H, I, J, K, 0, IF gain

 

 

accuracy, RBW switching, or log fidelity.

Log fidelity

Only

if

a

repair

is

made

to blocks D, F, H, K, IF gain accuracy, RBW

 

 

switching,

or log

fidelity.

If multiple adjustments are required they should be done in the following order:

1.Limiter phase

2.Linear fidelity

3.Log fidelity

 

TEST CABLE

 

----------------------,

:

I

S P E C T R U M I

I

I

SK16

Figure 2-9. DC Log Adjustment Setup

2-30 Adjustment Procedures