Table 1-4. Recommended Test Equipment (continued)

Instrument

Termination

1Low-pass filter

IJaw-pass filter

(two required)

IJaw-pass filter

(two required)

I‘ower splitter

siervice accessory kit 7Yuning tool

cZables 7Test cable

C:able, RG-214/U

C:able

C:able, 50 Sl coaxial

(jive required)

C!able

(two required)

Critical Specifications for Equipment Substitution

Frequency range: dc to 13.2 GHz

Impedance: 50 52

Maximum SWR: <l.lO

Connector: Type N(m)

Cutoff frequency: 50 MHz

Rejection at 65 MHz: >40 dB

Rejection at 75 MHz: >60 dB

Cutoff frequency: 1.8 GHz

Rejection at >3 GHz: >45 dB 0.1 dB ripple

Cutoff frequency: 4.4 GHz

Rejection at 5.5 GHz: >40 dB

Frequency range: dc to 13.2 GHz

Output tracking: <0.25 dB

insertion loss: 6 dB (nominal)

Equivalent output SWR: <1.22

No substitute

No substitute

Connectors: BNC(m)-to-SMB(f)

Length: 261 cm (24 in.)

Connectors: Type N(m)

Length: 291 cm (36 in.)

Connectors: SMA(m)

Length: 24 to 36 inches

?onnectors: BNC(m)

Length: 2 122 cm (48 in.)

rrequency range: 30 Hz to 26.5 GHz

tiaximum SWR: <1.4 at 26.5 GHz

vlaximum insertion loss: 3 dB

Yonuectors: APC 3.5(m), both ends

,ength: 2 61 cm (24 in.)

Recommendet M o d e l

HP 909A

0955-0306

0955-0491

HP 11689A

HP 11667B

08562-60021

8710-1010

85680-60093

HP 11500A

8126-1578

HP 10503A

8120-4921

U s e

P,M,V

P

P

A,T

A

AN

p,v

P

PAV

* Part of microwave workstation

P = performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting; V = operation verification