12.Figure 10-7 illustrates the waveforms in step 11 expanded to show relative timing. the second and fourth traces are delayed by 5 ms from the first and third. The oscilloscope settings are changed as follows:

Sweep time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 ps/div

13.Figure 10-8 illustrates the waveforms of properly working line generators. Whenever there is a pulse on TP2 (or TPl), the appropriate integrator U201B (or U203B) generates a ramp (the output vector) which feeds back to U201A (U203A) and shows on its output.

5 . 0 V/div 0 . 0 0 v

1 . 0 0 m s / d i v

0 . 0 0 0 s

U201A-1

TP2

U203A-1

TPI

SK195

Figure 10-6. Distorted X/Y Line Generator Waveforms

5 . 0 V/div 0 . 0 0 v

2 0 . 0 us/div 0 . 0 0 0 s

U201A-1

T P 2

U203A-1

TPI

SK196

Figure 10-7. Expanded X/Y Line Generator Waveforms

Controller Section 1 O-9

Page 366
Image 366