3.Use the up and down step keys to scroll through any other error messages which might exist, making note of each error code.

Error Message Elimination

When an error message is displayed, always perform the following procedure:

1.Press ISAVE) and SAVE STATE.

2.Store the current state in a convenient STATE register. (It may be necessary to set SA%LOCK to OFF.)

3.Press (CAL) and REALIGN LO & IF . Wait for the sequence to finish.

4.Press (z) and RECALL STATK .

5.Recall the previously stored STATE.

6.If an error message is still displayed, refer to the list of error messages below for an explanation of the error messages.

System Analyzer Programming Errors (100 to 150)

Refer to the HP 8560 E-Series Spectrum Analyzer User’s Guide for information on programming the spectrum analyzer.

100 NO PWRON Power-on state is invalid; default state is loaded. Press (SAVE),

PMR ON STATE to clear error message.

101 NO STATE State to be RECALLed not valid or not SAVEd.

106 ABORTED! Current operation is aborted; HP-IB parser reset.

107 HELLO ?? No HP-IB listener is present.

108 TIME OUT Analyzer timed out when acting as controller.

109 CtrlFail Analyzer unable to take control of the bus.

110 NOT CTRL Analyzer is not system controller.

111 # ARGMTS Command does not have enough arguments.

112 ??CMD?? Unrecognized command.

113 FREQ NO! Command cannot have frequency units.

114 TIME NO! Command cannot have time units.

115 AMPL NO! Command cannot have amplitude units.

116 ?UNITS?? Unrecognizable units.

117 NOP NUM Command cannot have numeric units.

118 NOP EP Enable parameter cannot be used.

119 NOP UPDN UP/DN are not valid arguments for this command.

7-16 General Troubleshooting

Page 237
Image 237
Agilent Technologies 856290216 service manual System Analyzer Programming Errors 100 to, Error Message Elimination