13. Demodulator Adjustment

8.A 1 kHz sine wave should be observed on the oscilloscope. Rotate the volume knob on the front panel of the spectrum analyzer until the amplitude of the 1 kHz signal is at about 150 mV (3 divisions on the oscilloscope).

9.Adjust A4C707 FM DEMOD for a maximum peak-to-peak response on the oscilloscope.

10. Press (LINE) to turn the spectrum analyzer off. Disconnect the test cable from A4C723.

CONNECT HERE

SP115E

Figure 2-22. Demodulator Adjustment Locations

Adjustment Procedures 2-55

Page 95
Image 95
Agilent Technologies 856290216 service manual Demodulator Adjustment Locations Adjustment Procedures