Access to Internal Assemblies

Servicing the HP 85623 requires the removal of the spectrum analyzer cover assembly and folding down six board assemblies. Four of these assemblies lay flat along the top of the spectrum analyzer and two lay flat along the bottom of the spectrum analyzer. All six assemblies are attached to the right side frame of the spectrum analyzer with hinges and fold out of the spectrum analyzer allowing access to all major assemblies. See Figure 4-l.

nTo remove the spectrum analyzer cover assembly, refer to procedure 1. w To access the A2, A3, A4, and A5 assemblies, refer to procedure 5.

w To access the Al4 and Al5 assemblies, refer to procedure 9.

nTo remove the Al6 or Al7 assembly, refer to procedure 10.