17. 16 MHz PLL Adjustment

8.Adjust A2R152 (16 MHz PLL ADJ) until the microwave frequency counter reads

14.4MHz 6200 KHz.

9.Reconnect W22 to A2J8. The microwave frequency counter should read 16 MHz. If the counter reads 16 MHz and the display is still distorted, perform the display adjustments in “Display Adjustment,” in this’chapter.

10.On the spectrum analyzer, press (CAL), REXICN LO arrd IF to clear any error messages.

Adjustment Procedures 2-65

Page 105
Image 105
Agilent Technologies 856290216 service manual MHz PLL Adjustment