13. Demodulator Adjustment

13. Demodulator Adjustment

Assembly Adjusted

A4 log amplifier/Cal oscillator assembly

Related Performance Test

There is no related performance test for this adjustment.

Description

A 5 kHz peak-deviation FM signal is applied to the INPUT 500. The detected audio is monitored by an oscilloscope. FM DEMOD is adjusted to peak the response displayed on the oscilloscope.

D I G I T I Z I N G

OSCILLOSCOPE

SIGNAL GENERATOR

A3

I OUTPUT

ADAPTERADAPTER

$B

sj150e

Figure 2-2 1. Demodulator Adjustment Setup

Equipment

AM/FM signal generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8640B Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501A

Adapters

Type N (m) to BNC (f) (2 required) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1476 Type N (f) to APC 3.5 (f) (Option 026 only) . . . . . . . . . . . . . . . . . . . . . . . . .1250-1745

Adjustment Procedures 2-53

Page 93
Image 93
Agilent Technologies 856290216 service manual Demodulator Adjustment