15.External Mixer Amplitude Adjustment (Non-Option 327)

16.YIG-Tuned Filter/Mixer (RYTHM) Adjustment

Assembly Adjusted

Al4 frequency control assembly

Related Performance Tests

Image, Multiple, and Out-of-Band Responses

Second Harmonic Distortion

Frequency Response

Description

The slope and offset of the A10 RYTHM tuning voltage are set by DACs on the Al4 frequency control assembly. The offset DAC value is optimized at a low frequency and the slope DAC value is optimized at a high frequency.

Automated Procedure Available

The YIG-Tuned Filter/Mixer adjustment is included in the automated HP 85623 “Front End Cal” adjustment, which is described in Chapter 3, “Frequency Response Adjustment Software.”

 

 

 

 

 

BNC CABLE

 

 

H‘R PROT/WR ENA

 

 

 

 

 

 

 

 

JUMPE :R IN

 

 

 

 

 

 

 

 

THIS AREA

 

 

 

 

‘EX

 

 

J

 

 

 

 

ADAPTER

 

 

II

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OUTPUT

Al4 FREP CONTROL

Al5 RF

Sk121

Figure 2-25. RYTHM Adjustment Setup

Equipment

Synthesized sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8340A/B

Adapters

Type N (m) to APC 3.5 (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1744 Type APC 3.5 (f) to APC 3.5 (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5061-5311

Cables

APC 3.5, 91 cm (36 in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8120-4921

2-60 Adjustment Procedures

Page 100
Image 100
Agilent Technologies 856290216 service manual YIG-Tuned Filter/Mixer Rythm Adjustment, Rythm Adjustment Setup