Agilent Technologies 856290216 service manual Blank Display Using the TAM

Models: 856290216

1 482
Download 482 pages 3.94 Kb
Page 455
Image 455

A 1 7

C R T D R I V E R

 

J 4

 

( R E V I S I O N

I

C O N N E C T O R )

I’

SK1121

Figure 13-2. Al7 Test Connector

Table 13-3. TAM Tests versus Test Connectors

 

 

Manual Probe Troubleshooting Test

 

Measured Signal Line!

 

 

 

 

 

Revision

 

 

 

MS5

 

 

Constant

current

Source

 

MS1

 

 

Intensity

input

 

 

MS7

 

 

Intensity

offset

 

 

MS7

 

 

Blanking

control

 

 

MS8

 

 

10 V reference test

 

MS4

 

Switch

drive test

 

 

MS3

 

Buffered X & Y DAC outputs

 

MS2, MS7

 

X

line

generator

test

 

MS6

 

Y

line

generator

test

 

MS1

 

Intensity

offset output

 

MS8

 

Revision

 

 

 

MS1

 

X, Y, & Z output offset

 

MS3, MS4, MS7

 

X

output

amplifier

 

MS7

 

Y

output

amplifier

 

MS3

 

Blanking

test

 

 

MS4

 

Focus DAC test

 

 

MS2

 

 

 

 

 

 

 

 

 

Blank Display (Using the TAM)

Use the following procedure if your spectrum analyzer has a blank display. This procedure substitutes an HP-IB printer for the display.

1.Connect the printer to the spectrum analyzer and set the printer address to the value required by the TAM (‘t1 is usually one). Do not connect any other controller to the spectrum analyzer.

Display/Power Supply Section 13-5

Page 455
Image 455
Agilent Technologies 856290216 Blank Display Using the TAM, Al7 Test Connector TAM Tests versus Test Connectors