10. Calibrator Amplitude Adjustment

10. Calibrator Amplitude Adjustment

Assembly Adjusted

Al5 RF assembly

Related Performance Test

Calibrator Amplitude and Frequency Accuracy

Description

The CAL OUTPUT amplitude is adjusted for -10.00 dBm measured directly at the front panel CAL OUTPUT connector.

YP C

BOARD

PROP

S K 1 1 3

Figure 2-18. Calibrator Amplitude Adjustment Setup

Equipment

Measuring receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8902A Power sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A

Adapters

Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1477

Procedure

Note

The spectrum analyzer

should be allowed to warm up for at least 30 minutes

 

before performing this

adjustment.

 

 

 

1.Place the spectrum analyzer in the service position shown in Figure 2-18. Prop the Al4 frequency control board assembly in the service position.

2.Zero and calibrate the HP 8902A/HP 8482A combination in log display mode. Enter the power sensor 300 MHz cal factor into the HP 8902A.

3.Connect the HP 8482A through an adapter directly to the spectrum analyzer CAL OUTPUT connector.

4.Adjust A15R561 CAL AMPTD for a -10.00 dBm reading on the HP 8902A display.

Adjustment Procedures 2-47