Procedure 14. W3 Line Switch Cable

20.With wire cutters, clip the tie wrap holding the cable to the contact housing, From the top side of the spectrum analyzer, use contact removal tool, HP Part Number 8710-1791, to remove the four wires from the W3 connector. See Figure 4-26.

21.Completely remove the cable from the instrument.

Note

If steps 20 and 21 have been performed, stop the procedure here; do not

 

perform steps 22, 23, and 24.

 

 

22.Remove the Al front frame assembly and Al8 CRT assembly as described in “Procedure 2. Al Front Frame/Al8 CRT Removal,” steps 16 through 29.

23.Remove the left side frame from the spectrum analyzer using the hardware listed below. (The side frame will still be attached by the speaker wires. Do not let it hang freely.)

Screw

Quantity

(1) SCREW-MACH M4 X 0.7 8 mm-LG FLAT HD . . . . . . . . . . . . . . . . . . . . . . .3

(2) SCREW-MACH M3 X 0.5 35 mm-LG FLAT HD . . . . . . . . . . . . . . . . . . . . . .2

(3) SCREW-MACH M3 X 0.5 6 mm-LG FLAT HD . . . . . . . . . . . . . . . . . . . . . . .6

24. Remove the line-switch cable assembly.

4-50 Assembly Replacement