Procedure 10.Al6 Fast ADC and Al7 CRT Driver
Removal
1. Remove the spectrum analyzer cover assembly and fold out the A2, A3, A4, and A5
assemblies as described in steps 3 through 6 under “Procedure 5. A2, A3, A4, and A5
Assemblies Removal.”
2. Place the spectrum analyzer top-side-up on the work bench with A2, A3, A4, and A5
folded out to the right.
3. For Option 007 spectrum analyzers: Remove two screws (1) securing the two
board-mounting posts to the left side frame, and remove the posts. See Figure 4-20.
4. Remove two screws (2) securing the Al7 assembly (and Al6 assembly in Option 007) to the
left side frame. Remove the two spacers (non-Option 007).
5. Pull the Al7 assembly out of the spectrum analyzer.
6. For Option 007 spectrum analyzers: Pull the Al6 assembly out of the spectrum analyzer.
7. Disconnect W7, W8,
W9,
A6AlW2,
and
A18Wl
from the Al7 CRT driver assembly.
8. For Option 007 spectrum analyzers: Disconnect all cables from the Al6 Fast ADC
assembly.
Figure
4-20.
Al6 and Al7 Mounting Screws
4-40 Assembly Replacement