Procedure 10. Al6 Fast ADC and Al7 CRT Driver

Removal

1.Remove the spectrum analyzer cover assembly and fold out the A2, A3, A4, and A5 assemblies as described in steps 3 through 6 under “Procedure 5. A2, A3, A4, and A5 Assemblies Removal.”

2.Place the spectrum analyzer top-side-up on the work bench with A2, A3, A4, and A5 folded out to the right.

3.For Option 007 spectrum analyzers: Remove two screws (1) securing the two board-mounting posts to the left side frame, and remove the posts. See Figure 4-20.

4.Remove two screws (2) securing the Al7 assembly (and Al6 assembly in Option 007) to the left side frame. Remove the two spacers (non-Option 007).

5.Pull the Al7 assembly out of the spectrum analyzer.

6.For Option 007 spectrum analyzers: Pull the Al6 assembly out of the spectrum analyzer.

7.Disconnect W7, W8, W9, A6AlW2, and A18Wl from the Al7 CRT driver assembly.

8.For Option 007 spectrum analyzers: Disconnect all cables from the Al6 Fast ADC assembly.

Figure 4-20. Al6 and Al7 Mounting Screws